TONAR Calibration: A Brief Note

  • Bernd Weinberg
  • J. Douglas Noll
  • Michael Donohue

Abstract

Measurements were made of the frequency response characteristics of the microphone-separator components of TONAR II instrumentation. The results of our calibration studies revealed 1) appreciable non-uniformity in frequency response of the two microphones, 2) a consid— erable degree of mismatch in frequency response between the microphones and, 3) dynamic interactions among microphone, separator cavity, and talker cavity resonant characteristics. Findings are discussed in terms of their implications regarding the validity of TONAR II based nasalance ratio measures.
Published
1979-04-01
Section
Articles